Search results for "Automatic extraction .CRM . PAHs . Sediments . GC-MS"
showing 1 items of 1 documents
Comparison of different methods for extraction of polycyclic aromatic hydrocarbons (PAHs) from Sicilian (Italy) coastal area sediments.
2012
This paper describes a work aimed at improving the conditions of an extraction method, coupling GC-MS determination, for the analysis without cleanup phase, of polycyclic aromatic hydrocarbons (PAHs) from sediment samples. The automatic Soxhlet extraction in warm mode (using Extraction System B-811 Standard, Buchi) has demonstrated advantages for automation, reduced extraction time, and lower solvent use than for conventional Soxhlet extraction. Under these conditions, the recoveries are very good as they resulted greater than 85 % and, in most of the cases, near 100 %. The repeatability is also satisfactory (relative standard deviation less than 15 %). The detection limits are also accepta…